Ul adapter factors, Lv models rv-ha-hb models – Brookfield UL Adapter User Manual
Page 4

Brookfield Engineering Laboratories Inc.
- 4 -
Operating Instructions M/91-080-I1003
The factor is used to calculate viscosity in units of centipoise (cP) or milli Pascal seconds (mPa•s) when using Dial
Reading or Model DV-I Viscometers. The Viscometer dial reading/display value x Factor = cP (mPa•s).
UL Adapter Factors
Figure 5: Pivot Cup Compatibility
TYPE I
TYPE II
TYPE III
Threaded
Hole
When an older dial reading visco-
meter is used, the viscometer Pivot
Cup may need to be replaced with
the current design. If your visco-
meter has a Type I or Type II pivot
cup, the cup should be replaced
with a Type III cup. Contact
Brookfield or your Brookfield agent
for information.
LV Models
RV-HA-HB Models
Shear
Shear
Factor
Speed
Rate
Speed
Rate
(RPM)
(sec
-1
)
Factor
(RPM)
(sec
-1
)
RV
HA
HB
60.0
73.38
0.1
100.0
122.30
0.64
1.28
5.12
30.0
36.69
0.2
50.0
61.15
1.28
2.56
10.24
12.0
14.68
0.5
20.0
24.46
3.20
6.40
25.60
6.0
7.34
1.0
10.0
12.23
6.40
12.80
3.0
3.67
2.0
5.0
6.12
12.80
25.60
1.5
1.83
4.0
4.0
4.89
16.00
0.6
0.73
10.0
2.5
3.06
25.60
0.3
0.37
20.0
2.0
2.45
1.0
1.22
0.5
0.61
Shear Rate = RPM x 1.223
SPECIALISTS IN THE
M E A S U R E M E N T A N D
CONTROL OF VISCOSITY
T
EL
508-946-6200 or 800-628-8139
F
AX
508-946-6262
w w w. b ro o k f i e l d e n g i n e e r i n g . c o m
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346-1031 USA