Appendix b — high temperature calibration fluids – Brookfield Thermosel with EZ-Lock User Manual
Page 17

Brookfield Engineering Laboratories, Inc.
Page
Manual No. M09-208
APPENDIX B — hIGh TEMPERATURE CALIBRATION FLUIDS
These fluids are recommended for verifying the calibration of the Brookfield Thermosel System.
Fluid
Temperature
°C
°F
Nominal Viscosity
Centipoise (mPa•s)
HT - 30,000
25.0
77
30,000
93.3
200
9,000
149.0
300
4,500
HT-60,000
25.0
77
60,000
93.3
200
18,000
149.0
300
9,000
HT-100,000
25.0
77
100,000
93.3
200
30,000
149.0
300
15,000
The viscometer calibration is verified first. A regular* "disc" spindle should be used to test the HT
fluid in its jar at 25.0°C or 77°F. Please refer to the appropriate viscometer operation manual for
instructions.
The Thermosel System and its appropriate spindle are then used to measure the HT fluid's viscosity
at 200 and 300°F.
Contact Brookfield Engineering Laboratories or an authorized dealer/distributor/representative in
your area for purchase of these fluids.
* Standard spindles supplied with viscometer.