beautypg.com

Din* spindle ranges for small sample adapter, Table 3 ) – Brookfield Small Sample Adapter with EZ-Lock User Manual

Page 4

background image

Brookfield Engineering Laboratories, Inc.

- 4 -

Manual No. M09-091-A0912

(Table 3)

DIN* Spindle Ranges for Small Sample Adapter

MODEL

VISCOSITY RANGE (cP)

Spindle 82

Spindle 83

Minimum

-

Maximum

Minimum

- Maximum

LVT

5.7

10,000

18.9

37,898

LVDV-I+

3.5

10,000

11.3

37,898

LVDV-II+

1.7

10,000

5.7

50,000

LVDV-III

1.4

10,000

4.5

50,000

RVT

36.5

10,000

121.0

50,000

RVDV-I+

37.5

10,000

121.0

50,000

RVDV-II+

18.7

10,000

60.5

50,000

RVDV-III

15.0

10,000

48.3

50,000

HAT

75.0

10,000

242.0

50,000

HADV-I+

75.0

10,000

242.0

50,000

HADV-II+

37.5

10,000

121.0

50,000

HADV-III

29.2

10,000

97.0

50,000

HBT

300

10,000

967.2

50,000

HBDV-I+

300

10,000

967.2

50,000

HBDV-II+

150

10,000

483.6

50,000

HBDV-III

120

10,000

387

50,000

SMC value

3.75

12.13

SRC value

1.29

1.29

Spindle Entry Code

82

83

Sample Volume (mL)

5.5

1.5

*DIN spindles conform to DIN 53019.
The 82 spindle (SC4-DIN-82) works in an SC4-13R or SC4-13RP chamber.