DIN* Spindle Ranges for Small Sample Adapter
MODEL
VISCOSITY RANGE (cP)
Spindle 82
Spindle 83
Minimum
-
Maximum
Minimum
- Maximum
LVT
5.7
10,000
18.9
37,898
LVDV-I+
3.5
10,000
11.3
37,898
LVDV-II+
1.7
10,000
5.7
50,000
LVDV-III
1.4
10,000
4.5
50,000
RVT
36.5
10,000
121.0
50,000
RVDV-I+
37.5
10,000
121.0
50,000
RVDV-II+
18.7
10,000
60.5
50,000
RVDV-III
15.0
10,000
48.3
50,000
HAT
75.0
10,000
242.0
50,000
HADV-I+
75.0
10,000
242.0
50,000
HADV-II+
37.5
10,000
121.0
50,000
HADV-III
29.2
10,000
97.0
50,000
HBT
300
10,000
967.2
50,000
HBDV-I+
300
10,000
967.2
50,000
HBDV-II+
150
10,000
483.6
50,000
HBDV-III
120
10,000
387
50,000
SMC value
3.75
12.13
SRC value
1.29
1.29
Spindle Entry Code
82
83
Sample Volume (mL)
5.5
1.5
*DIN spindles conform to DIN 53019.
The 82 spindle (SC4-DIN-82) works in an SC4-13R or SC4-13RP chamber.
The 83 spindle (SC4-DIN-83) works in an SC4-7R or SC4-7RP chamber.