E0608

Brookfield Engineering Laboratories, Inc.

Page 20

Manual No. M00-151-H0612

7) The viscosity reading should equal the cP value on the viscosity standard fluid to within the

combined accuracies of the Viscometer and the viscosity standard (as discussed in the section

entitled, Interpretation of Calibration Test Results) which appears later in this section.

Verification Procedure for a Small Sample Adapter

When a Small Sample Adapter is used, the water jacket should be connected to a water bath and the

sample stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies

with each spindle/chamber combination. (Refer to the Small Sample Adapter instructions).

2) Place the sample chamber into the water jacket.

3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the Viscometer.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.

5) Measure the viscosity and record the Viscometer reading.

Note: The spindle must rotate at least five (5) times before a viscosity reading is taken.

6) The viscosity reading should equal the cP value on the viscosity standard fluid to within the

combined accuracies of the Viscometer and the viscosity standard (as discussed in the section

entitled, Interpretation of Calibration Test Results) which appears later in this section.

Verification Procedure for a Thermosel System

Before verifying the Thermosel and controller, follow the verification procedure for regular spindles

RV/HA/HB 2-6 or LV 61-63+. When a Thermosel System is used, the controller stabilizes the Thermo

Container at the test temperature.

1) Put the proper amount of HT viscosity standard fluid into the HT-2 sample chamber. The

amount varies with the spindle used. (Refer to the Thermosel instruction manual).

2) Place the sample chamber into the Thermo Container.

3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the Viscometer.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.

5) Measure the viscosity and record the Viscometer reading.

Note: The spindle must rotate at least five (5) times before a viscosity reading is taken.

6) The viscosity reading should equal the cP value on the viscosity standard fluid to within the

combined accuracies of the Viscometer and the viscosity standard (as discussed in the section

entitled, Interpretation of Calibration Test Results) which appears later in this section.