beautypg.com

Ii. getting started – Brookfield Dial Stepper Viscometer User Manual

Page 10

background image

E0608

Brookfield Engineering Laboratories, Inc.

Page 10

Manual No. M00-151-H0612

II. GETTING STARTED

II.1 Operation

MOTOR

ON/OFF/PAUSE

POWER SUPPLY

RECEPTACLE

SPEED KNOB

(rpm)

CLUTCH LEVER

Note: See Appendix D for
image of Explosion Proof
Viscometer.

GUARD LEG

CLAMP

1. Be sure the plug of the power supply

is securely plugged into your power

source.

2. Plug the metal jack of the power

supply into the circular receptacle

on the back of the Viscometer.

3. The black, rubber baton switch on

the viscometer controls the motor

and has three positions:

UP

: Off - turns the motor off and

stops the dial from turning

MIDDLE

: On - causes the dial to

rotate at the selected speed

DOWN

: Pause (or motor stop)

- causes the dial to pause when

rotating

4. LV Viscometers use a set of four spindles and a narrow guard leg; RV Viscome ters use a set

of six spindles and a wider guard leg; HA and HB Viscometers use a set of six spindles and no

guard leg.

5. Speeds (rpm) are changed by turning the black knob on the top of the viscometer (to the left or

right) to the desired speed.

6. The clutch lever, when depressed, raises the dial against the red pointer and "holds" the Viscometer

reading. Releasing the clutch lowers the dial and frees the pointer.

Top View

Front View