beautypg.com

Troubleshooting – Wilbur Curtis TCC1N User Manual

Page 7

background image

7

TROUBLESHOOTING

SYMPTOM

POSSIBLE CAUSE

SOLUTION

Difficulty priming the TCC tea

dispenser.

Adjusting rod on mixing valve is turned

in too far.

Open up the valve 1/4 turn counter

clockwise.

Tea does not dispense from

faucet.

Water line is turned off.

Turn on water line valve.

Only water flowing from fau-

cet, no tea.

1. Tea concentrate BIB is out of product.
2. The tea BIB line is kinked.

3. The tea BIB line is clogged.

4. The check valve in the tea concen-

trate line may be stuck closed.

5. The adjustment screw may be turned

in or closed.

1. Replace the tea concentrate BIB.
2. Trace the tea line and straighten out

kinked hose.

3. Blow out hose to clean blockage or

replace with a new hose.

4. Loosen the clamps and remove the

check valve. Clean or replace it.

5. Open the mixing valve to resume

product flow.

Faucet is dripping constantly.

1. The faucet seat cup may have hard-

ened particles keeping the faucet

from sealing.

2. There may be a tear or cut in the

seat cup.

3. Water pressure may be too high.

1. Disassemble and clean the seat cup

(see cleaning section).

2. Replace the seat cup.

3. Call CURTIS CARE for instructions

on adjusting the water regulator.

Finished tea is weak.

1. The mixing valve is out of adjust-

ment.

2. The product check valve may be

stuck.

1. Open the mixing valve to increase tea

concentrate flow.

2. Remove and clean the check valve.

Make sure check valve is installed

with correct flow direction.

Product flowing from the fau-

cet is too foamy or bubbly.

1. The clamps at the check valve may

not be tight.

2. The clamp on the QCD connector

may be loose.

3. Tea concentrate product may have

air in the BIB.

4. The O-ring on the QCD is worn out.

1. Reconnect tubing and tighten clamps.

2. Make sure the barbed end of the

QCD fitting is pushed in on the hose

and the clamp is tight.

3. Turn the BIB so the fitting faces up-

ward and open the faucet until all the

bubbles are released.

4. Replace the worn O-ring or install a

new QCD.

This manual is related to the following products: