Thr thr on – Torbal BA Platform Series User Manual
Page 49
- 48 -
Threshold check weighing is used to check whether a weight of an object falls within specified threshold
limits.
Function Options:
Thr Off
(throFF) Disables threshold and check weighing.
Thr On
(thr on)
Enables threshold weighing and sets function parameters.
Thr Prn
(thrPrn)
Prints currently set parameters.
Dat Cfg
(thrCFG)
Sets output mode for the opto-isolator (Pulse Mode / Level Mode).
23.1. Enabling check weighing and setting threshold limits.
1. Press the MENU key, when the option “thr” is displayed press the T key, and then select the option
“thr On” by press the T key once again.
2. The scale will begin to display threshold limits. To set the Lower Limit wait for “Set-1” to display and
press the T key. To set the Upper Limit wait for “Set-2” to display and press the T key. To set the
minimum weight value that should be used in check weighing select “Set-3”.
3. The scale will display a dashed line indicating to enter the limit value. To enter the value, use the
following keys: the C key to increment a digit, the T key to accept and go to the next digit, the
key to
insert a decimal, and the MENU key to accept the entire setting.
To increment a digit
To accept and go to
next digit
To insert a decimal
To accept entered
value
Chapter 23: Threshold Check Weighing and Opto-Isolator (ThR)
thr
thr on
set 0
set 1
set 2
set 3