After a weighing transaction is completed, a result data receipt can be printed. To initiate printing, press the
data transfer key. Data may be sent to a printer or a PC via the Torbal Communication Software.
15.1. Data Transmission and Exchange Protocol
Data Transmission (LONG):
Transmission Parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps,
Exchange data:
Transmit the weight (equivalent to the Print Key,
, in weighing:
Computer
→Scale: S I CR LF (53h 49h 0Dh 0Ah) – initiating signal,
Scale
→Computer: scale sends 16 Bytes of data as follows:
Byte
1
- The character ‘-‘ or space
Byte 2 -
space
Bytes
3,4 - digit or space
Bytes
5-9 - digit, comma, or space
Byte 10
-
digit
Byte 11
-
space
Byte
12 - k, l, c, p or space (for kg,lb,ct,pc, or%)
Byte
13 - g, b, t, c or %
Byte 14
-
space
Byte 15
-
CR
Byte 16
-
LF
‘Tare the weight’ (corresponds to the
→T← key in weighing):
Computer
→Scale: S T CR LF (53h 54h 0Dh 0Ah),
Scale
→Computer: no response.
‘Zero the scale’ (corresponds to the key
→0← in weighing):
Computer
→Scale: S Z CR LF (53h 5Ah 0Dh 0Ah),
Scale
→Computer: no reponse.
‘Turn On / Off the Scale (corresponds to the key I/ in weighing):
Computer
→Scale: S S CR LF (53h 53h 0Dh 0Ah),
Scale
→Computer: no response.
‘Display the MENU’ (corresponds to the key MENU in weighing):
Computer
→Scale: S F CR LF (53h 46h 0Dh 0Ah),
Scale
→Computer: no response.
Setting the threshold 1 (optional):
Computer
→Scale: S L D1...DN CR LF (53h 4Ch D1...DN 0Dh 0Ah)
where: D1...DN – Threshold value, up to 8 characters,
Chapter 15: Port 1 - RS232 Communication Port Configuration (Port-1)