beautypg.com

Specifications, Chapter 2: specifications – Torbal AGN100 User Manual

Page 6

background image

- 5 -

Model


AG100

AG200

AG300

AG500

AG1000

AG2000

AG3000

AG4000

Capacity

100g

200g

300g

500g

1000g

2000g

3000g

4000g

Readability (d)

0.001g

0.01g

Repeatability (Standard

Divination)

0.001g

0.01g

Linearity

+/-0.002g

+/-0.02g

Stabilization time (typical)

Approx. 3 sec

Approx. 3 sec

Tare Range

-100g

-200g

-300

-500

-1000

-2000

-3000

-4000

Accuracy Class

I

Automatic Internal

Calibration

YES

Time Configuration

1h, 2,h, 3h, 4h, 5h, 6h

Temperature

Configuration

0.5°C, 1.0°C, 1.5°C, 2.0°C

Automatic Calibration

Adjustment Masses (g)

20,50, 100

50, 100,

200

100, 200,

300

100, 200,

500

200, 500,

1000

500, 100,

2000

1000,
2000,

3000

1000, 2000,

4000

Pan Dimension

120 mm

165 mm x 165 mm

Draft Shield

Optional

Not Required

Scale Dimension

216 x 345 x 90 mm

Operating Temperature

+15°C to +30°C

PS2 Port

Standard

RS232 Port

Bidirectional

USB

Standard

Power Supply

Input: 120VAC 60Hz 15W Output: 12VDC 500mA

Display Type

Graphical LCD

Display Size

30mm x 100mm

Application Modes

Weighing, Parts Counting, Animal Weighing, Check Weighing, Display Hold, Percent Weighing, Totalizing, Check

Counting, Compounding, Statistics, SQC Quality Control, 512 KB Library Bank

Weighing Units

Grams, Milligrams, Kilograms, Carats, Pounds, Newton, Grains, Ounces Troy, Pennyweight

Scale’s Net Weight (lb/kg)

5kg/11lbs

Sample Size

1 through 100 or any

Parts

Counting

Min Piece

Weigh

30mg

300mg

Sample Size

1 through 100 or any

Check

Counting

Min Piece

Weigh

30mg

300mg

Min

Reference

Weight

100mg

1 g

% Resolution

Displayed

Percent

Weighing

(Reference

Weight)

0.1% (100mg to 7g), 0.01% (7g to 70g)

0.001% (>70g)

0.1% (1g to 70g), 0.01% (70g to 700g)

0.001% (>700g)

Tiny Intervals

3sec., 5sec., 10sec., 15sec.

Animal

Weighing

Modes (Min

Weight)

Manual, Auto (10mg), Semi-Auto (10mg)

Manual, Auto (100mg), Semi-Auto (100mg)

Totalizing

Modes (Min

Weight)

Manual, Auto (10mg), Semi-Auto (10mg)

Manual, Auto (100mg), Semi-Auto (100mg)

Display

Hold

Modes (Min

Weight)

Manual, Auto (10mg), Semi-Auto (10mg)

Manual, Auto (100mg), Semi-Auto (100mg)

Chapter 2: Specifications

This manual is related to the following products: