Lipo termination voltage 4.20v, Life termination voltage 3.60v delay 0min, Lilo termination voltage 4.10v – ProgressiveRC 106B+ iCharger User Manual
Page 12: Pb termination voltage 2.40v
-12-
iCharger Synchronous Balance Charger/Discharger 106B+
LiPo Termination
Voltage 4.20V
LiFe Termination
Voltage 3.60V
Delay 0min
Lilo Termination
Voltage 4.10V
-
Dec
+
Inc
-
Dec
+
Inc
-
Dec
+
Inc
Pb Termination
Voltage 2.40V
-
Dec
+
Inc
Inc
Dec
Inc
Dec
Inc
Dec
Termination voltage for LiFe. Adjust LiFe‘s charge termination voltage.
Press Start/Enter for more than 3 seconds to blink the ―Voltage‖ value,
then press Dec and Inc to adjust the termination voltage.
Range: 3.40V – 3.90V, Step: 0.01V (3.60V, default)
Termination voltage for Lilo. Adjust Lilo‘s charge termination voltage.
Press Start/Enter for more than 3 seconds to blink the ―Voltage‖ value,
then press Dec and Inc to adjust the termination voltage.
Range: 3.90V – 4.20V, Step: 0.01V (4.10V, default)
Termination voltage for LiPo. Adjust LiPo‘s charge termination voltage.
Press Start/Enter for more than 3 seconds to blink the ―Voltage‖ value,
then press Dec and Inc to adjust the termination voltage.
Range: 4.00V – 4.30V, Step: 0.01V (4.20V, default)
Termination voltage for Pb. Adjust Pb‘s charge termination voltage. Press
Start/Enter for more than 3 seconds to blink the ―Voltage‖ value, then
press Dec and Inc to adjust the termination voltage.
Range: 2.20V – 2.50V, Step: 0.01V (2.40V, default)
Note: The Termination voltage controls the transition from Constant
Current (CC) to Constant Voltage (CV) charging for LiXx and Pb
chemistries. It also defines the per-cell limit for cell overvoltage detection.
If you change the default termination voltage, the charge and discharge
setting screen will note this difference by alternatively blinking battery‘s
type and setting voltage value.
Inc
Dec