HI-4455 Isotropic Electric Field Probe
© ETS-LINDGREN L.P. – August 2005
Revision D- Part No. H-600076
32
PROBE OUTPUT
Command
Description
B
Bxx.xx, where xx.xx is the battery voltage.
D1
Dxx.xxuuu, the short form output.
xx.xx is the reading. The position of the decimal point
depends upon the range setting of the probe.
uuu = units
_V_ = V/m, mW2 = mW/cm², _V2 = [V/m]²
(underscore indicates a space character)
D2
Dxx.xxuuurrrobaaat, the long form output.
xx.xx = the reading, as described for D1.
uuu = units, as describe for D1.
rrr = recorder out value (A 3-digit ASCII number from 0
to 255).
o = over range indicator ("N" = ok, "O" = over range).
b = battery status ("N" = safe operating level, "W" =
warning level, "F" = fail level).
aaa = axis enable ("E" = enabled).
t = terminating carriage return.
Rx
Rx, where x is the range.
x = "" returns the range currently in use
x = 1, 2, 3, 4 enables the selected range
x = N sets the probe to the next (higher) range.
TF
Txxx, where xxx is temperature in E Fahrenheit.
TC
Txxx, where xxx is temperature in E Centigrade.
Vtx
Vxx[xx...], where xx is table data in ASCII hex.