Page — 14

HI

n

3550 Manual

The HI

n

3550 can display integrated magnetic flux levels

of up to 999.99 T-h. Integrated flux levels in excess of

this value are not displayed.

5.1 Using the Integrating (

II

B) Mode

1. Slide the AC-DC (Magnetic Field Selection) switch to

select the type of magnetic field to be measured.

Set the STBY-

I

B-

)

B (Measurement Mode) Switch to

I

B.

NOTE

Do not perform the next step if you wish to add

to previously stored data.

2. Press the SELECT and SET buttons (CLEAR function)

on the back of the instrument simultaneously for

approximately three seconds to clear the memory of

stored integrated time and value.

The alarm settings remain unchanged after the memory

is cleared.

Changing the Measurement Mode

Changing from integrating to instantaneous measurement

mode will stop the integrating measurement mode while

retaining the integrated time and value. After you change

the measurement mode, the reading will remain

unchanged until one measurement cycle (3 seconds) is

completed.