Archived 4/2/10 – ETS-Lindgren HI-3550 Magnetic Field Monitor (Archived) User Manual

Page 18

background image

HI

n

3550 Manual

Page — 13

Figure 5

Integrated Time and Integrated Value Cleared

5.0 DESCRIPTION OF HI

n

n

3550 OPERATION

Introduction

This section provides a detailed description of the

operation of the HI

n

3550.

Measurement Cycle

The HI

n

3550 performs a measurement every three

seconds and displays the results on the lower line of the

display panel.

Measurement Ranges

The HI

n

3550 can measure magnetic field strength up to

0.3 T for each of the three axes. Depending on the

orientation of each sensing axis with respect to the

direction of the magnetic field, it is possible to display the

measured value of field intensities greater than 0.3 T.

In the integrating mode, the HI

n

3550 displays elapsed

time on the upper line of the display panel. "H" denotes

hours, "M" denotes minutes, and "S" denotes seconds.

Integrating time is displayed up to 999 hours, 59

minutes, 59 seconds. When elapsed time exceeds this

value, the display rolls over to 0 hours, 00 minutes, 00

seconds, and restarts the count.

Archived 4/2/10