Functional Description

MODEL 4630 REFRAD REFERENCE RADIATOR

12

© EMC TEST SYSTEMS, L.P. – MARCH 2002

REV F – PN 399214

correction factors is the worst – case chamber factor. The mean
between the upper and lower envelope curve is the mean chamber
factor.

Chamber factors (in dB) are to be added to the field-strengths (in
dB

μV/m) measured in the camber in order to achieve open-site

equivalents (in dB

μV/m).


During operation with CalStan, the REFRAD, test receiver or
spectrum analyzer, and antenna positioning mast are remotely
controlled. The operator is guided through the procedure by
commands appearing on the screen.

FUNCTIONAL DESCRIPTION

COMB GENERATOR


Functional groups:

Remote control decoder

Crystal

oscillator

and

divider

Switching

power

supply

Pulse

generator

Matching

network


The clock signal from the crystal oscillator is divided to the
selected pulse repetition rate. With this signal the pulse generator
produces nano-second pulses with a very high amplitude, requiring
a voltage of 70 V generatored by the switching power supply. A
resistive 50 Ohm matching network is provide between the pulse
generator and the output.

Due to the use of a regulated switching power supply, the output
amplitude is stable for a 4.4 to 6 V power supply range.

REMOTE CONTROL UNIT


Functional groups:

GPIB – bus decoder

Remote

control

signal

encoder

Selection

and

display

unit