Remote control – ETS-Lindgren 4630 Refrad Reference Radiator (Archived) User Manual
Page 10

Technical Data
MODEL 4630 REFRAD REFERENCE RADIATOR
6
© EMC TEST SYSTEMS, L.P. – MARCH 2002
REV F – PN 399214
REMOTE CONTROL
POWER SUPPLY 
Plug in power supply, 9V DC, 50 mA. 
 
GPIB BUS 
Selection of address 
The GPIB address (DEFAULT) is changed by 
removing the top of the remote control unit and 
selecting the desired switch settings on the GPIB 
interface boards (see diagram below). The table relates 
the switch setting to the actual address. (Note that 
moving the switch up turns it on and pushing it down 
turns it off.) 
 
GPIB COMMAND SET 
The GPIB commands are listed in the following table: 
LINE SPACING
GPIB 
COMMAND 
Standby “0” 
0.01 MHz 
“1”
1 MHz
“2”
5 MHz
“3”
NOTE: The commands have to sent without 
termination character(s) and without “CRIF”. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
FIGURE 5. Remote Control Unit System Diagram.
GPIB
ADDR
A5 A4 A3 A2 A1
0
OFF OFF OFF OFF OFF
1
OFF OFF OFF OFF ON
2
OFF OFF OFF ON OFF
3
OFF OFF OFF ON ON
4
OFF OFF ON OFF OFF
5
OFF OFF ON OFF ON
6
OFF OFF ON ON OFF
7 OFF
OFF ON ON ON
8
OFF ON OFF OFF OFF
9
OFF ON OFF OFF ON
10 OFF ON OFF ON OFF 
11 OFF ON OFF ON ON 
12 OFF ON ON OFF OFF 
13 OFF ON ON OFF ON 
14 OFF ON ON ON OFF 
15 OFF ON ON ON ON 
16 
ON OFF OFF OFF OFF
17
ON OFF OFF OFF ON
18
ON OFF OFF ON OFF
19
ON OFF OFF ON ON
20
ON OFF ON OFF OFF
21
ON OFF ON OFF ON
22
ON OFF ON ON OFF
23
ON OFF ON ON ON
24
ON ON OFF OFF OFF
25
ON ON OFF OFF ON
26
ON ON OFF ON OFF
27
ON ON OFF ON ON
28 ON ON ON OFF
OFF
29 ON ON ON OFF ON 
30 ON ON ON ON OFF 
31 ON ON ON ON ON 
Address - Switch setting table