Series H-491198-36 Battery Charger

Environmental

Operating Temperature:

10

°

C–40

°

C (50

°

F–104

°

F)

Humidity:

5% to 95% relative humidity,

non-condensing

M

AINTENANCE

R

ECOMMENDATIONS

Operate the battery charger with care.

There are no user serviceable parts inside the battery charger.

Opening the battery charger housing may void your warranty.

R

EPLACING THE

F

USE

Disconnect the battery charger from power

before replacing a fuse.

If the battery charger fails to operate, check for a blown fuse inside the power

entry module. A blown fuse must be replaced with the same value and type of

fuse, or an unsafe condition may result. Use only 250 Volt, 1.0 Amp, Type T

(5 mm x 20 mm) fuses.

To replace a fuse:

1.

Two fuses are located in the fuse drawer in the power input module.

Use a screwdriver to open the drawer.

2.

The fuse towards the outside of the drawer is the spare. Remove the

spare fuse from the module.