Operation

| 21

3.

Plug the charger jack completely into the CHARGE connector on the

clamp-on current sensor.

Allow approximately one hour for a full charge cycle. When charging is

complete, the battery charger automatically goes into a trickle charge.

It will remain in trickle charge mode until disconnected from the

HI-3702.

4.

When the charge cycle is complete, remove the charger jack from the

CHARGE connector on the clamp-on current sensor.

5.

Unplug the battery charger from the AC outlet.

6.

Use the HI-4416 to verify that the battery is charged to approximately

3.8 VDC.

See Series H-491198-36 Battery Charger on page 27 for complete information

on the battery charger.

Error Codes

If an error occurs, the probe will respond with one of the following strings. These

strings begin with a colon and end with a carriage return.

E01

Communication error (for example, overflow)

E02

Buffer full error; too many characters contained between the

start character and carriage return sequence

E03

Received command is invalid

E04

Received parameter is invalid

E05

Hardware error (for example, EEPROM failure)

E06

Parity error

E9

Received command is invalid