10

|

Maintenance

Battery Maintenance

Under normal operating conditions, recharging the battery is the only

HI-3702 maintenance performed by the user.

Maintaining the HI-3702 requires that the nickel-cadmium (NiCd) battery be

properly charged and the battery condition monitored. The normal operational life

of the battery (approximately 1000 charge/discharge cycles) is such that by the

time the battery needs replacement; the HI-3702 will need recalibration; these

tasks are performed by ETS-Lindgren. For information on calibration services,

see Annual Calibration on page 10.

For the steps to charge the battery, see page 20.

In case of damage to the battery that causes premature failure, the HI-3702 must

be sent to ETS-Lindgren for battery replacement; replacement in the field is not

possible. See Service Procedures on page 11 for information on returning the

HI-3702 to ETS-Lindgren for service.

Annual Calibration

See the Product Information Bulletin included with your shipment for information

on ETS-Lindgren calibration services.

Maintenance of Fiber Optics

Fiber optic connectors and cables can be damaged from airborne particles,

humidity and moisture, oils from the human body, and debris from the connectors

they plug into. Always handle connectors and cables with care, using the

following guidelines.