4.0 BATTERY CHARGING

Introduction

Each HI

n

3637 meter contains a rechargeable nickel-

cadmium (NiCd) battery. A fully-charged battery (nominal

output voltage of 12 VDC) provides up to 20 hours of

operation.

NOTE:

Holaday Industries, Inc., charges the internal

NiCd battery of the HI

n

3637 at the factory in

order to calibrate the instrument prior to

shipment. While every effort is made to

ensure that your meter arrives ready to use,

we cannot guarantee that this will be the

case. Always check the condition of the

meter's battery prior to use.

Battery Tips

Nickel-Cadmium (NiCd) batteries have several

characteristics that can affect both their performance and

operating life. The following tips advise you how to take

advantage of these characteristics to get the most out of

your meter's battery.

!

Although NiCd batteries are rated for operation in

temperatures from -20 °C to +65 °C (-4 °F to

+140 °F), using the meter in extreme temperatures

will reduce operating time significantly. The

optimum operating temperature range for these

batteries is +20 °C to +30 °C (+68 °F to +86

°F).

!

The battery in the HI

n

3637 does not require

periodic "deep discharges" to reverse the capacity-

depleting "memory effect" caused by repeated

shallow discharges; however, undercharging can

reduce battery capacity.

!

If the battery appears unable to acquire or maintain

an appreciable charge, individual cells in the battery

may be shorted or damaged. If, for any reason, you

need assistance replacing your battery contact

Holaday Customer Service.