beautypg.com

ETS-Lindgren HI-2602 Interlock Monitor User Manual

Page 8

background image

8

|

Introduction

The diode detection array of eight hot carrier diodes is housed in the large end of

the plastic probe. This antenna array has the unique ability to sum microwave

electric fields of any polarization in a plane perpendicular to the axis of the probe.

The probe is attached to the main unit by a BNC connector. The main unit and

the probe are calibrated as an integral unit. Temporary over-exposure of up to

2000 mW/cm² will not cause probe burn-out or damage to the instrument or its

calibration.

Standard Configuration

HI-2602 main unit

Probe cable assembly (including probe sensor cover)

Amp plug and hood

Amp pins, 9 pieces

• User

manual

ETS-Lindgren Product Information Bulletin

See the ETS-Lindgren Product Information Bulletin included with your shipment

for the following:

• Warranty

information

Safety, regulatory, and other product marking information

Steps to receive your shipment

Steps to return a component for service

ETS-Lindgren calibration service

ETS-Lindgren contact information