Menu Options, Settings, Command Set
31
AUDIO
Turns on the audible alarm indicator, which will sound to indicate that
the measured leakage has exceeded a preset level (alarm set point).
Although the alarm set point may be specified, the alarm will
not sound unless the audio is turned on. For set point
information, see page 30.
Menu Option
Setting
Serial Port
Command
OFF
Q0
ON
Q1
—
Q?
AUTO-RANGE
When set to auto-range, the best range for each measuring condition
will automatically be selected.
Menu Option
Setting
Serial Port
Command
Function
—
A0
OFF
—
A1
ON
—
A?
Retrieve status
The auto-range setting only affects the bar graph display
and the analog output.
When in Peak Hold with the range set to AUTO, the range
will change as the field strength increases; it will remain in
the range that shows the highest peak value until the ZERO
button is pressed. This is so the peak value can be
displayed on the bar graph. When the ZERO button is
pressed with no field present, the range will revert back to
the lowest (1 mW/cm2).