Alarm setpoint, E 30 – ETS-Lindgren HI-1710A Microwave Oven Survey Meter User Manual
Page 30
![background image](/manuals/636272/30/background.png)
30
Menu Options, Settings, Command Set
www.ets-lindgren.com
ALARM
SETPOINT
Sets a level after which an alarm will sound to indicate that the
measured leakage has exceeded the level.
The alarm will sound only if the audio is turned on. To turn
on the alarm, see Audio on page 31.
Menu Option
Setting
Serial Port
Command
XX.XX
T####
—
T?
Where #### is a value from 0 to 9999 in mW/cm
2
. For
example, 5000 is 5.0 mW/cm
2
, 5 is 0.005 mW/cm
2
, and so
on.
Each press of the up/down arrow button will raise/lower the setting
by 0.01 for the first ten presses (9.99, 9.98, 9.97, and so on), by 0.10
for the next ten presses (9.89, 9.79, 9.69, and so on), and by 1.00 for
the remaining presses (8.89, 7.89, 6.89, and so on).
Any time a change in direction is made up or down, the change
increment begins at 0.01.
The alarm will sound only when the displayed RF value exceeds the
alarm set point. For example, if the alarm is set at 4.99 mW/cm
2
, the
alarm will only sound when the displayed RF level is 5.00 or greater.
Once the RF level decreases below the alarm set point, the alarm will
turn off.