Operation
25
Starting the HI-1710A
1.
With the power off, attach the HI-2623 Microwave Oven Probe to the
probe connector on the front of the HI-1710A. Insert the connector so
that the locating screw is facing upward.
2.
Connect the HI-1710A to AC power.
3.
Turn on the power switch on the front of the HI-1710A.
The screen will display the firmware version and additional
ETS-Lindgren information. During this time a series of self-tests are
performed; when the self-tests are complete, the probe is ready to be
zeroed.
4.
Place the probe in a zero field environment, and then press ZERO on
the front of the HI-1710A.
When zeroing is complete, the HI-1710A is ready for operation.
Using the HI-1710A
Check the probe spacer cone for damage, discoloration, or
contamination. Accuracy may be affected by cone wear or damage or
by contamination by metallic particles which may become imbedded in
the cone. If necessary, replace the cone. See Replacement and
Optional Parts on page 15 for part numbers.
1.
Allow the HI-1710A to warm up for at least 10 minutes.
2.
Select the operating settings for the test needs. For normal tests or a
beginning point, use the default conditions. To select settings, see
Menu Options, Settings, and Command Set on page 27.
3.
Zero the probe before beginning the test. With the oven operating with
a suitable load, move the probe across the oven surfaces, keeping the
probe axis (handle) perpendicular to the potential leak area.
Hold the probe by the red handle to avoid potential interference. Keep
other parts of the body and the probe cord away from the immediate
vicinity of the probe head.