beautypg.com

ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual

Page 24

background image

Page — 16

HI

n

1501 Manual

Preparing the Meter

1.

Remove the meter from the carrying case. Make

sure the EPS (styrofoam) spacer cone is intact and

firmly seated on the probe. Check the cone for

signs of wear or contamination. Replace if worn or

contaminated.

2.

Move the SLOW/FAST switch to the “FAST”

position. Turn the selector switch to “BATT TEST.”

The needle should indicate above the green “Batt

OK” line. If not, replace both batteries (refer back

to page 8).

3.

Turn the selector switch to the “PROBE TEST”

position. The needle should indicate between the

green “OK”, “PROBE TEST” lines. If not, the probe

or meter circuits have been damaged and should not

be used. A defective probe cannot be replaced in

the field and the complete unit should be returned

to the factory for repair.

4.

Turn the selector switch to the desired scale

(usually the 2 mW scale) and let the meter stabilize

for two minutes.

5.

Adjust the indicator needle on the meter to read

zero by turning the ZERO ADJ knob. The probe

must be shielded or removed from the vicinity of

any RF fields when the meter is being zeroed.