ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual

Page 14

background image

Page — 6

HI

n

1501 Manual

Parameter

Error (dB)

1 mW/cm

2

Calibration - Precision

- Accuracy

+0.42, -0.16

±0.09

Nonlinearity and AM response

±0.09

Near Field vs. Far Field

±0.29

Receiving Pattern

-0.11

Temperature Response

±0.17

Frequency Response *

±0.04

Polarization *

±0.21

RFI *

±0.04

Drift *

±0.02

Total

+1.27 / -1.12

* Errors combined in RMS manner

Calibration Method

Each meter is calibrated by placing the probe in a CW

(continuous wave) 2450 MHZ electromagnetic field

generated by a crystal controlled solid state source. The

source feeds an anechoic chamber through a section of

waveguide terminating in a slot radiator. Calibration is

performed at a level of 1 mW/cm

2

and a standard

temperature of 75

/

F.

The calibration is also checked at 5 mW/cm

2

. The

accuracy of the field is determined by comparing with an

LCR (local calibration reference) which is traceable to

NIST (National Institute of Standards and Technology