2.0

ABOUT THE HI-1501

Description

Microwave leakage (electromagnetic fields) is detected by

an array of eight hot carrier diodes housed in the large

end of the plastic probe. This antenna array has the

unique feature of being able to sum microwave electric

fields of any polarization in a plane perpendicular to the

axis of the probe. The antenna lobe (effective measuring

area) is also very broad, making the instrument easy to

use when measuring leakage around an oven door. The

spacer cone is designed to provide a 5 cm spacing from

the tip of the probe to the center of the array.

Each detection probe and meter amplifier is calibrated as

a unit and serves all three ranges: 2, 10, and 100

mW/cm

2

. Only the 2 and 10 mW/cm

2

ranges are

considered compliance ranges in terms of compliance

with the DHHS/CDRH standard for microwave survey

instruments contained in CFR 1030.10.

NOTE:

The HI-1501 must not be used without the spacer cone

in place. Using the instrument without the cone will

result in reading errors and may damage the unit.