ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual

Page 15

background image

HI

n

1501 Manual

Page — 7

[formerly NBS]) through FDA/CDRH (Food and Drug

Administration / Center for Devices and Radiological

Health).

Maintenance

The only maintenance that is required is the replacement

of the batteries or the spacer cone, should it become

damaged or worn. Both items may be purchased from

Holaday Industries, Inc.

NOTE:

Do not use the HI-1501 without the spacer in position.

To maintain stated instrument accuracy, use only spacer

cones provided by Holaday.

Battery Replacement

Turn the selector switch OFF. Unsnap the two clips

that hold the cover in place. Always replace both

batteries at the same time. Any nine (9) volt alkaline

(or the optional, rechargeable NiCad) battery may be

used. Replace batteries and reinsert the meter

assembly into the cover being careful to not crimp the

battery wires between the front panel and the cover.

After replacement, turn the selector switch to BATT

TEST. If the meter does not immediately indicate full

scale, check the condition of each battery to assure

that they are in good condition or fully charged.

Spacer Cone Replacement

The spacer cone must be replaced when worn or

contaminated. Simply remove the old cone by pulling

straight off the end of the probe. Slip a new cone on,

being sure the cone is fully seated on the probe. A

polystyrene cone shield is available which provides

protection for the relatively soft material of the cone.