beautypg.com

ETS-Lindgren HI-1501 Microwave Oven Survey Meter User Manual

Page 13

background image

HI

n

1501 Manual

Page — 5

Accuracy

The accuracy of this instrument is within ± 1 db

(+25%, -20%) when used according to these

instructions:

1. Always use a clean Holaday spacer cone.

Accuracy will be affected by wear of the cone and

by dirt and metallic particles which may become

imbedded in the EPS (expanded polystyrene -

Styrofoam).

2. If the instrument is to be used continuously for an

extended period of time, the condition of the

batteries must be checked by rotating the selector

switch to the BATT TEST position at least once

every 15 minutes.

3. The normal operating temperature range is

between 60

/

and 90

/

F. If used outside this

range, an additional error will be introduced. This

error can be approximated as -0.14% per

Fahrenheit degree. The negative temperature

coefficient means that the instrument reads high

at lower temperatures and low at higher

temperatures.

4. The parameters which affect the accuracy of this

instrument are listed along with the error

contribution of each in the following table.