beautypg.com

Inspection, Asme b107.14m, Iso6789 – CDI Torque TORQUE SCREWDRIVERS User Manual

Page 7

background image

Page 4

INSPECTION

Inspect all wrenches repaired to determine that they are in an operational condition by

visually inspecting for worn, broken, cracked or damaged parts. Replace parts not found

to be in satisfactory condition.

ACCURACY VERIFICATION AND MINOR CALIBRATION PROCEDURE

This procedure is to be used to check the accuracy and, if necessary, perform a minor

adjustment on calibration. It must also be used after a major calibration.

There are two major international standards that establish torque wrench accuracy

requirements. They are ASME B107.14 Hand Torque Tools (American Society of

Mechanical Engineers) and ISO 6789 Hand Torque Tools – Requirements and Test

Methods (International Organization for Standardization). Both standards are recognized

as established accuracy test methods and referenced in this section. ASME and ISO

standards are subject to review and changes. The repair manual will be reviewed and

revised periodically but may not reflect the latest revisions of the specifications listed

above.

ASME B107.14

1. Preload the wrench by setting the wrench at 100% of full scale and operating three

times on the tester.

2. Reset handle to 20% of full scale. Take 3 readings at 20%.

3. Set and test the wrench three (3) times at 60% and 100% of full scale noting the

accuracy error for each reading. If the readings are within the accuracy requirements

for the direction (±6% or readings), no calibration adjustment is required.

ISO 6789

1. Preload the wrench by setting the wrench at 100% of full scale and operating five (5)

times in the direction of the test.

2. Reset handle to 20% of full scale.

3. Test the wrench five (5) times at 20% noting the accuracy error for each reading. Set

and test the wrench five (5) times at 60% and 100% of full scale noting the accuracy

error for each reading. If the readings are within the accuracy requirements for the

test direction (±6% or readings), no calibration adjustment is required.