3. Place the 1/8” T-Shaped Hex in the back of the screwdriver and Turn the set screw to
set the calibration as follows:
c.) If the reading on the tester was below the accuracy limit for the value tested turn
the set screw clockwise to increase the value. Test and repeat this procedure until
the readings are within the accuracy limits.
2. If the reading on the tester was above the accuracy limit for the value tested turn the
set screw counterclockwise to decrease the value. Test and repeat this procedure
until the readings are within the accuracy limits.
3. Place the end plug on the end of the screwdriver.