Calibration

All SP Series scales may be calibrated with any precision test weight from 10 lb to 100 lb. The recom-

mended test weight is about 2/3 of the full scale capacity, but in general, the larger the test weight, the

more accurate the scale will be. The only restriction is that the test weight

must

be in pounds.

To calibrate the scale:

Turn the scale off. Locate the hidden switch on the base panel and set it at the rightmost position.

Turn the unit back on. The message "

C 0

" appears on the display briefly, followed by a value which

remains on the screen. Allow a 20 minute warm-up period for the load cell and indicator to become

thermally stable.
Press

ZERO

to zero the value, then press the

NET/GROSS

key to save the zero point value.

The display will momentarily prompt "

C 1

" for the span calibration, followed by "

0.00

" with one digit

flashing. Place the test weight on the platform.
Use the four directional keys to adjust the displayed value to the actual test weight value in pounds.

Increase the flashing digit by pressing the

lb/kg

key. Decrease the flashing digit by pressing the

ZERO

key. The position of the flashing digit may be changed by pressing the

PRINT

key or the

TARE

key.

After setting the exact value, press the

NET/GROSS

key to save the value.

If the calibration was successful, the display will show "

ECAL

" momentarily, then freeze. Exit the

Calibration mode and enter the Normal Operating Mode by turning off the scale, positioning the

calibration switch back to the leftmost position, and turning the scale back on.
If the calibration was

not

successful, one of the error messages below will appear. Take the indi-

cated action to correct the problem, then perform a new calibration.

"Err0"

- The calibration test weight or the adjusted keyed-in weight is larger than full scale. Change

the calibration test weight or check the keyed-in weight.

"Err1"

- The calibration test weight or the adjusted keyed-in weight is smaller than 10% of full scale.

Change the calibration test weight or check the keyed-in weight.

"Err2"

- Check keyed-in weight with the actual weight placed on platform.