beautypg.com

4 calibration procedure, 1 required equipment, 2 connections – Dynalco DST-2000C Direction Sensing Tachometer User Manual

Page 9: 1 required, Equipment

background image

DST-2000C Instruction Manual

9

2.3

Wiring Connections Between DST-2000C & M343 Pickup Connector

Wiring connections between the DST-2000C and the M343 magnetic pickup
connector C917-X (X = length of connector cable in inches) are as follows:

2.4 Calibration

Procedure

The only calibration to the DST-2000C recommended by the factory is calibration of
the 0–1 mA Meter Output. {Note: The two potentiometers on the bottom of the PC
board are factory adjustments only and no user adjustment should attempted.}

2.4.1 Required

Equipment

Dynalco F-16 Frequency Generator or similar square wave generator

Dynalco G-101 meter or digital voltmeter

Dynalco isolation transformer p/n 260C-6495

2.4.2 Connections

Connect the G-101 to the meter output terminals of the DST-2000C: terminals
23 (+) and 24 (–).

Connect the F-16 signal generator to the DST-2000C terminals 9 (+) and 10
(–).

Connect the red lead of the 260C-6495 transformer to terminal 9 of the
DST-2000C and the yellow lead to terminal 10.

Connect the white lead of the 260C-6495 transformer to terminal 12 and the
black lead to terminal 13 of the DST-2000C.