The Performance Series centrifuges are capable of storing up to 10 user-defined presets. These memory
presets contain all of the settings needed to define a specific run, (run time, speed, acceleration rate,
braking rate, etc.). The user can use these memory locations to quickly configure the centrifuge for a
specific test type and ensure that the test is run in the same way each time by recalling this same setting.

Store a configuration in a memory location

1. Adjust the various settings (speed, time, acceleration rate, etc.) for the

configuration to be saved.

2. Press and hold the PROGRAM button until the preset display begins to flash.

3. Press the MEMORY button to increment the memory location by one. Continue

to press the memory button until the desired location is displayed.

4. Press the program button to store the current configuration into the memory

location that is displayed.

Recall a configuration from a memory location

1. Press the MEMORY button until the desired memory location is displayed in the

memory display.

2. As you review the presets, the settings on the control panel will change to

reflect the settings stored in the memory preset currently displayed.

3. When the desired preset is found, pause and the memory display will

momentarily flash. The stored configuration is now loaded.

Notes:

Once a configuration has been loaded, it will remain active until any change is made on the

control panel. If for instance the user adjusts the time, the current configuration will no longer match
the previously loaded configuration and the memory display will change to a “-” dash. The previous
configuration may be recalled using the procedure described above.

Memory locations may be overwritten by selecting them during the store procedure.

To display the current cycle count, perform the following:

1. Unlock the lid by pressing the OPEN/STOP button.

2. Open the lid latch so that the yellow LATCHED indicator light is not illuminated.

3. Press and hold the START button for approximately five seconds. The word “CYCLES” will

This centrifuge is equipped with a digital cycle counter to monitor machine usage and for routine
maintenance.

Memory Locations / Storing and Recalling

Digital Cycle Counter

be displayed in the speed display and the current cycle count will be displayed in the time
display. If the current cycle count exceeds 9,999 cycles, the cycle count will be displayed
by showing the first part of the number and then the end of the number. For example: the
1st shown is 0002 and the 2nd shown is 4537 represents 24,537 cycles.