Section 4 |
Operating Instructions
Opening and Closing the Lid
The plate may be hot. Use caution and check temperature on display before handling slides.
Improper precaution can cause a burn.
Depressions located on either side of the lid allow user to lift lid into position. The lid should offer
some resistance when opening. To close reverse process, assure front is completely down and no
obstructions prevent gasket seal from sealing on housing base.
Turn Unit On
The Hybridizer main power switch is located on the rear panel. Assure unit is plugged into the appropriate
outlet. Move switch to “ON” position. Instrument will beep to announce power has been turned on, fan
and heating will start and the Main Menu will be displayed when the instrument has reached the default
temperature of 37°C.
I = ON O=OFF
Run a Denat & Hyb Program
Turn unit on and wait for the Main Menu screen. Cursor highlights “Run a PGM” line.
Press “Enter” button to accept.
With the arrow keys scroll through program numbers 1 to 40 / program names. If no programs have been
saved advance to programming section of this handbook. To accept, press “Enter” button.
Run a PGM
Edit a PGM
Create a PGM
Present Temp: 37°C
Enter PGM no.
or Scroll (arrows)
PGM 01 HER2