Dako Hybridizer User Manual
Page 14

14
Dako Hybridizer
|
Handbook
If the ambient temperature changes during a process and causes the instrument process set temperature 
to change beyond the +/- 1
°C specification for more than 2 minutes, a message will appear to inform the
user that the ambient temperature is “high”. A constant beep will sound. The counter will continue to 
count. The present temperature will be displayed. Hitting the “Stop” button will allow user to abort the 
process. A new screen will be displayed asking user if they are sure they want to abort. 
 
Hyb only: 
 
“Ambient Temp High!” flashes
alternately on “Hyb in Process” line.
 
 
 
Fixed Temp: 
 
“Ambient Temp High!” flashes
alternately on “Fixed Temp” line.
 
 
 
Abort Screen Message: 
 
 
 
 
PGM -- Name 
Hyb in Process 
Hyb --
°C --:--
Present Temp: --
°C
PGM – Name
Fixed Temp --
°C
Reset Timer 00:00:00 
End PGM/Main Menu
ABORTING!!
Are You Sure? 
No 
Yes – Main Menu