Dako Hybridizer User Manual
Page 14

14
Dako Hybridizer
|
Handbook
If the ambient temperature changes during a process and causes the instrument process set temperature
to change beyond the +/- 1
°C specification for more than 2 minutes, a message will appear to inform the
user that the ambient temperature is “high”. A constant beep will sound. The counter will continue to
count. The present temperature will be displayed. Hitting the “Stop” button will allow user to abort the
process. A new screen will be displayed asking user if they are sure they want to abort.
Hyb only:
“Ambient Temp High!” flashes
alternately on “Hyb in Process” line.
Fixed Temp:
“Ambient Temp High!” flashes
alternately on “Fixed Temp” line.
Abort Screen Message:
PGM -- Name
Hyb in Process
Hyb --
°C --:--
Present Temp: --
°C
PGM – Name
Fixed Temp --
°C
Reset Timer 00:00:00
End PGM/Main Menu
ABORTING!!
Are You Sure?
No
Yes – Main Menu