Introduction:
This manual covers the installation and operation of Humboldt FlexPanels
for Triaxial and Permeability testing. This equipment comes in different
configurations but is essentially built from two basic units. The equipment
is packaged for easy installation, which allows you to be up and running as
quickly as possible.
The primary intention of this manual is to guide the first-time user of FlexPanels
through the familiarization and installation of the equipment. After the initial
installation is accomplished the manual can serve as an ongoing reference for
installation changes or future reference on the equipment.
While this manual is intended to include as much available information as
possible, it is designed for those who will not be reading the entire manual or
who only want to refer to specific sections. Users should classify themselves and
proceed accordingly:
First-Time User: We recommend that all first time users and those basically
unfamiliar with FlexPanel operation read the entire manual before proceeding
with installation.
Experienced User: Refer to Precautions and Maintenance.
Unpacking and Inspection
Carefully remove the contents of the carton in which the panel was shipped.
Inspect the carton and the panel and make note of any apparent physical
damage. If severe damage is present then you should consider rejecting the
shipment and contacting the shipping company concerning in-transit damage
claims. We have made every effort at the factory before shipping to fully
inspect, test, and properly package this product so that it reaches you
defect-free and without damage.
All packaging materials should be saved and set aside in case a return shipment
needs to be made.
Precautions and Warnings
This equipment was designed to work with a maximum of 150 PSI (1000 kpa)
of air pressure. If pressures greater than 150 PSI (1000 kpa) are used it could
be dangerous for the operator, as well as cause damage to the equipment.
Leaks in the lines which contain water will influence the test results. Leaks
can occur at tube connections especially those which are taken apart and
reconnected. The quick-connects will leak if the o-ring has been cut, which
will happen when the o-ring is not lubricated. Parker makes an o-ring lubricate
called Super-O-Lube designed to be used with O-ring seals.