beautypg.com

Humboldt HM-4160 FlexPanels User Manual

Page 23

background image

Back Pressure Saturate Sample
The following steps should be followed to back pressure saturate the sample.
1) Turn the Selection Valve (A2) to Pressure.
2) Turn the Bias Valve to Bias On.
3) Turn the Burette Valve (V2) to Burette.
4) Turn the Selection Valve (A3) to off.
5) Turn the Bridge Valve (BV) to Bridge On.
6) Turn the Burette Valve (V3) to Burette.
7) Make sure the Top and Base Valves on the Triaxial cell, which are connected

to the FlexPanel, are open. Check the cell pressure by lifting the Toggle

Valve (T1).

8) Check the level of the water in the Burettes in Section 2 (Base) and 3 (Top)

to make sure they are over half full.

9) Apply back pressure to Sample.

a) Take Pore Pressure reading and Burette readings for Section 2 (Base)

and 3 (Top).

b) Lift to open Toggle Valve (T2) in Section 2.

c) Adjust the back pressure using Regulator (R2). Raise the pressure to

5 psi (35 kpa). When the pressure is increased, check the sample and burette

to make sure the Membrane is not blowing up. If water from the Burettes in

Section 2 (Base) and 3 (Top) start filling the membrane, close the Top and

Base Valves on the Triaxial Cell and check pressure in Section 1 (Cell) to

make sure it has increased. The pressure can be read on the Digital Display

on the Control Panel.

d) Check cell pressure by lifting Toggle Valve (T1) in section 1 and read the

pressure on the Digital Display on the control panel. The cell pressure

should have increased by 5 psi (35 kpa). If the pressure did not increase by

5 psi (35 kpa)

adjust the Pressure Regulator (R1) to the appropriate reading.

The following steps should be followed to check B valve

during back pressure

saturation.
1) Check the back pressure, lift Toggle Valve (T2) in section 2 (Base). Make

adjustments in reading if necessary.

2) Check the cell pressure, lift Toggle Valve (T1) in section 1 (Cell) and make

adjustment in reading if necessary.

3) Close the Top and Base Drainage Valves on the base of the Triaxial Cell.
4) Take Pore Pressure Transducer reading. The pore pressure transducer is

part of the Triaxial cell when we do the triaxial cell manual will address the pore

pressure transducer.

This manual is related to the following products: