beautypg.com

Humboldt HM-4160 FlexPanels User Manual

Page 19

background image

The pressure should be checked and adjusted after approximately 15 minutes

and then checked periodically when taking readings on from the Burettes.
Toggle Valve
The Toggle Valves are all connected in a series of tees, leading to the Digital

Display on the Control Panel. When the Toggle Valve is lifted, the pressure

for the adjacent Control Burette is indicated on the Digital Display. Only one

Control Burette pressure can be read at one time.
Bias Valve
Burette Section 2 (Base) is the only section with a Bias Valve (BA). This valve

is located below the Section 2 Toggle Valve. The Bias Valve (BA) is used to

combine the output pressure of Section 2 (Base) with that of Section 1 (Cell).
This is accomplished through the use of an extra port in the Regulator for

Section 1, which allows it to connect the output from Section 2 (Base) via the

Bias Valve. When the Bias Valve is open, it will connect the output of Section 2

with Section 1. Therefore, when adjustments are made to the Section 2 (Base)

Regulator (R2) the Section 1(Cell) Regulator (R1) will be changed by the same

amount. This feature can be used during back pressure saturation, allowing the

operator to use the Section 2 regulator (R2) to alter pressures, which maintains

the pressure difference between Section 1 (Cell) and Section 2 (Base).
Bridge Valve
Section 3 (Top) is the only Section with a Bridge Valve. This valve is located

below the Section 3 Toggle Valve. The Bridge Valve, when open, connects the

output of the Section 2 Regulator, to the top of the Section 3 (Top) Control

Burette. The Section 3 Pressure Selection Valve (A3) should be in the off position

when using the Bridge. If the Bias and Bridge Valves are both open, the Section

2 Regulator can be used to increase the pressure equally to all three Control

Burette Sections.
Burette and External Valves
These valves are located at the bottom of the controls in each Control Burette

Section. The Burette Valve is connected to the bottom of the Internal Burette

of the Burette Assembly, while the External Valve is connected to the Outer

Burette. Opening one or both of these valves connects them with the Quick

Connect in Section 1 (Cell). When the Burette Valve is open, the Internal Burette

is connected to the Section 1 Quick Connect. When the External Valve is open

the External Burette is connected to the Section 1 Quick Connect. These Valves

can both be open at the same time, which results in the Internal and the External

Burette seeking the same water level. When monitoring volume changes, only

the Burette Valve should be open.

This manual is related to the following products: