beautypg.com

Humboldt HM-4160 FlexPanels User Manual

Page 15

background image

Control Panel Connections (HM-4140, HM-4150, HM-4160):
(1/4-inch tubing is used for connections)
1) Install a section of tubing from the air supply to the PRESS IN connector on

the lower service connector plate.

2) Install a section of tubing from the water line to the TAP connector on the

lower service connector plate.

3) Install a section of tubing from the water side of the de-aired water

tank to the DEAIRED connector, marked de-aired on the lower service

connector plate.

4) Install a section of tubing from the vacuum supply to the VAC connector on

the lower service connector plate.

5) Install a section of tubing from a drain pipe or tank to the DRAIN

connector on the lower service connector plate.

6) Install a section of tubing from the top or air side of the de-aired water tank

to the AIR connector on the upper service connector plate.

7) Install a section of tubing from the water side of the de-aired water tank to

the WATER connector on the upper service connector plate.

Add-On Panel Connections (HM-4150A, HM-4160A):
(1/4-inch tubing is used for connections)
1. Install a section of tubing between the two service connectors of the

Control Panel PRESS OUT and the Add-On Panel PRESS.

2. Install a section of tubing between the two service connectors of the

Control Panel READOUT and the Add-On Panel READOUT.

3. Install a section of tubing with UNION TEE FITTING (HM-4150.25)

between the two service connectors of the Control Panel DEAIRED

and the Add-On Panel DEAIRED.

4. Install a section of tubing with UNION TEE FITTING (HM-4150.25)

between the two service connectors of the Control Panel DRAINAGE

and the Add-On Panel DRAINAGE.

5. Install a section of tubing with UNION TEE FITTING (HM-4150.25)

between the two service connectors of the Control Panel VAC (vacuum)

and the Add-On Panel VAC.