Humboldt HM-4160 FlexPanels User Manual
Page 15

Control Panel Connections (HM-4140, HM-4150, HM-4160):
(1/4-inch tubing is used for connections)
1) Install a section of tubing from the air supply to the PRESS IN connector on
the lower service connector plate.
2) Install a section of tubing from the water line to the TAP connector on the
lower service connector plate.
3) Install a section of tubing from the water side of the de-aired water
tank to the DEAIRED connector, marked de-aired on the lower service
connector plate.
4) Install a section of tubing from the vacuum supply to the VAC connector on
the lower service connector plate.
5) Install a section of tubing from a drain pipe or tank to the DRAIN
connector on the lower service connector plate.
6) Install a section of tubing from the top or air side of the de-aired water tank
to the AIR connector on the upper service connector plate.
7) Install a section of tubing from the water side of the de-aired water tank to
the WATER connector on the upper service connector plate.
Add-On Panel Connections (HM-4150A, HM-4160A):
(1/4-inch tubing is used for connections)
1. Install a section of tubing between the two service connectors of the
Control Panel PRESS OUT and the Add-On Panel PRESS.
2. Install a section of tubing between the two service connectors of the
Control Panel READOUT and the Add-On Panel READOUT.
3. Install a section of tubing with UNION TEE FITTING (HM-4150.25)
between the two service connectors of the Control Panel DEAIRED
and the Add-On Panel DEAIRED.
4. Install a section of tubing with UNION TEE FITTING (HM-4150.25)
between the two service connectors of the Control Panel DRAINAGE
and the Add-On Panel DRAINAGE.
5. Install a section of tubing with UNION TEE FITTING (HM-4150.25)
between the two service connectors of the Control Panel VAC (vacuum)
and the Add-On Panel VAC.