beautypg.com

Bio-Rad Media Sampler Pack User Manual

Page 6

background image

4

Fig. 1. A simulated chromatography profile from which HETP and As values are
calculated.

a b

10% of peak

h i ht

W

h,

width at half heig

Bed height = 20 cm
Column Volume = 20 L
Flow rate = 100 cm/h
Ve = 10 L
W

h

= 1 L

HETP = 0.036 cm
a = 1
b = 1.2
A

s

= 1.2

Bed height = 20 cm
Column volume = 20 L
Flow rate = 100 cm/h
V

e

= 10 L

W

½h

= 1 L

HETP = 0.036 cm
a = 1
b = 1.2
A

s

= 1.2

W

½h

width at half height

10% of peak

a

b

4110109C.qxd 11/1/2004 11:19 AM Page 5