5.3 Application 2: Removal of Affi-Gel
®
or Affi-Prep
®
MAPS
®
II Binding Buffer
from Antibody Solutions
1. Follow steps 1 through 3 in the general desalting
procedure. (Note: The equilibration buffer should
contain a minimum of 10 mM NaCI for IgG
monoclonal antibodies, or 50 mM NaCI for IgM
monoclonal antibodies.)
2. Load up to 3.0 ml of the elution fraction collected
from the Affi-Gel or Affi-Prep protein A column.
3. Add 8.0 ml of a suitable buffer to elute the antibody.
Collect 1.0 ml fractions.
4. Analyze the fractions to determine which fractions
to pool.
Section 6
Storage Conditions
Econo-Pac 10DG columns should be stored at room
temperature in 0.02% sodium azide.