System components, Unpacking the system components – Bio-Rad ZOE™ Fluorescent Cell Imager User Manual
Page 12
4 |
ZOE Fluorescent Cell Imager Instruction Manual
4 |
Table 1. ZOE Fluorescent Cell Imager specifications. (cont.)
FOV
0.70 mm
2
Camera
Monochrome, 12 bit CMOS, 5 megapixels
Image format
JPEG, TIFF, RAW (8 and 12-bit)
Internal memory
16 GB
Focus
Coarse and fine, manual adjustment
Stand-alone instrument
Yes
LCD monitor
10.1", color, LCD, touch screen with antiglare and antifingerprint treatment,
1,280 x 800 pixels image resolution, 80–180° angle tilt range
Ports
2 USB ports
Dimensions
33 x 32 x 30 cm (13 x 12.6 x 11.6")
Weight
9 kg (19.7 lb)
* Brightfield channel is illuminated with green LEDs that improve image contrast by reducing chromatic aberration.
Photobleaching due to this green light source is comparable to that encountered with a regular white light.
Note: User interface controls for individual channels are labeled according to the emission colors (blue, green, and red).
System Components
Catalog #145-0031 includes:
n
ZOE Fluorescent Cell Imager
n
Power cord
n
USB key
n
Instruction manual
n
Quick guide
Unpacking the System Components
1. Unpack the ZOE Cell Imager carefully. Remove all packaging materials and store them for
future use. Examine the instrument carefully for any damage incurred during transit. Ensure
that all parts of the instrument listed above are included with the product. If any item is
missing or damaged, contact your local Bio-Rad office.
2. Place the imager in an upright position on a dry, level surface.
3. Thanks to its light shield, the ZOE Cell Imager can be used in ambient light; dark room
operation is not required. However, it is not designed to be operated in direct sunlight.
4. Insert the supplied power cord into the instrument with its flat side facing down and without
blocking the locking jacket.
Note: To remove, pull the locking jacket to unlock the cable. Pulling the cable without unlocking
this mechanism can result in instrument breakage.
5. Plug the power cord into the appropriate electrical outlet and the instrument will power on.
The loading screen displays. After the loading sequence is completed the Position mode will
appear onscreen.
Note: Turning the unit off by pressing the Power off button results in only partial shutoff. For
complete power off remove power cord from electrical outlet.