Page 21

4.5 RNase Decontamination
Sub-Cell parts can be cleaned with a mild detergent and treated for 10 minutes with 3%
hydrogen peroxide (H
2
O
2
) and then rinsed with 0.1% DEPC- (diethyl pyrocarbonate) treated
distilled water to eliminate RNases prior to using the Sub-Cell systems for RNA gels
1-2
. Do
not soak Sub-Cell parts in DEPC water. Consult references
1-2
for other suggestions regarding
the use of DEPC in RNase decontamination.
Caution: DEPC is a suspected carcinogen. Always wear gloves, eye glasses and a
laboratory coat. Use caution when handling DEPC-containing solutions. Consult the
DEPC MSDS for more information.
Do not attempt to RNase decontaminate Sub-Cell parts using extreme dry heat.
Note: Several commercial products are also available for eliminating RNase contamination.
RNaseZAP
™
(Ambion) or RNase AWAY
™
(Molecular Bio-Products) are safe, simple and
effective methods that if used properly do not craze or fog the Sub-Cell parts. See
manufacturer instructions for proper use.