Symptoms
Cause
Solutions
Sample wells cast
Comb should be placed 1 to 2 mm
through the gel. Sample
above the base of the running surface.
leaks along bottom of
running surface.
Sample overload
Dilute sample.
Bands sharp but too
Gel agarose percentage too high
Lower agarose percentage.
few bands seen
Incomplete digestion
Check enzyme activity, digest further.
High MW bands
Gel agarose percentage too low
Increase agarose percentage.
sharp; Low MW
Switch to polyacrylamide.
bands smeared
Gels crack
Too high voltage gradient,
Reduce voltage. Run gel at lower
especially with low melting
temperature.
temperature agarose or
low gel strength gels
Section 6
Product Information
6.1 Sub-Cell
®
GT Systems
Catalog
Number
Product Description
170-4401
Sub-Cell GT System, with 15 x 10 cm tray
170-4402
Sub-Cell GT System, with 15 x 15 cm tray
170-4403
Sub-Cell GT System, with 15 x 20 cm tray
170-4404
Sub-Cell GT System, with 15 x 25 cm tray
170-4481
Sub-Cell GT System, with 15 x 10 cm tray and gel caster
170-4482
Sub-Cell GT System, with 15 x 15 cm tray and gel caster
170-4483
Sub-Cell GT System, with 15 x 20 cm tray and gel caster
170-4484
Sub-Cell GT System, with 15 x 25 cm tray and gel caster
170-4405
Wide Mini-Sub Cell GT System, with 15 x 7 cm tray
170-4485
Wide Mini-Sub Cell GT System, with 15 x 7 cm tray and mini-gel caster
170-4406
Mini-Sub Cell GT System, with 7 x 7 cm tray
170-4486
Mini-Sub Cell GT System, with 7 x 7 cm tray and mini-gel caster
Sub-Cell Systems with PowerPac
™
Basic Power Supply, 100—120/220—240 V
164-0302
Sub-Cell GT Cell and PowerPac Basic Power Supply
164-0301
Wide Mini-Sub Cell GT Cell and PowerPac Basic Power Supply
164-0300
Mini-Sub Cell GT Cell and PowerPac Basic Power Supply
14