Specifications
Trans-Blot®
Turbo™ Instrument
Dimensions: (L x W x H) 26.0 x 21.1 x 20.4 cm
Weight: 8 lb with cassettes, 4.5 lb without cassettes
Input power: 100–240 VAC, 276 VA, 50–60 Hz, 175 W max
Fuses: Two 6.3 A, 250 V, fuses located above the power connection
On/off switch: Yes, mains connected
USB port: Yes, input only, for firmware updates
Cooling fan: Yes
Output power: 0–26 VAC (1 V increments), 0–2.6 A DC (0.1 A increments) for each cassette
Operating conditions: 15–31°C ambient temperature, 0–95% relative humidity (noncondensing)
User Interface
18 button keypad, 128 x 64 pixel monochrome display
Programmable methods: Up to 25 user defined
Preprogrammed methods: Standard SD, 1.5 mm gels, High MW, Low MW, Mixed MW, 1 Mini-PROTEAN®
TGX™ gel
Audible alarm: Yes
User notifications:
• Power fail during run
• No load detection
• No cassette detection
• End of run
• Watt/hr limit
Trans-Blot Turbo Cassettes
Dimensions: (L x W x H) 20.2 x 16.0 x 4.5 cm
Anode: Platinized titanium electrode plate
Cathode: Stainless steel
Weight: 1.5 lb
Gel Compatibility
Suitable for transfer of two mini format gels or one midi fomat gel per cassette. Other gel sizes can be
trimmed to fit the consumable transfer packs.