beautypg.com

Bio-Rad Mini-PROTEAN® Tetra Handcast Systems User Manual

Page 9

background image

Sample calculation: preparing twelve 1.0 mm 4–20% gradient gels:

Casting twelve 1.0 mm gels requires 100 ml; prepare 110 ml.

Divide the total volume by 2 to get the volume required for each chamber. (For this example,
make 55 ml for the light chamber and 55 ml for the heavy chamber.)

Solution Volume Calculations:
Light Solution (4%)
Acrylamide
30% stock solution
(30%) (X ml) = (4%) (55 ml)

X = 7.3 ml

Tris-Cl Buffer
1.5M Tris-Cl stock buffer pH 8.8
(1.5M) (X ml) = (.375M) (55 ml)

X = 13.8 ml

Water
(55 ml)–(7.3 ml + 13.8 ml) = X

X = 34 ml

APS
10% Stock solution
(500 µl) / (100 ml) = (X µl) / (55 ml)

X = 275 µl

TEMED
10% of the APS volume; (275 µl) / 10 = X

X = 27.5 µl

Heavy Solution (20%)
Acrylamide
30% Stock solution
(30%) (X ml) = (20%) (55 ml)

X = 36.7 ml

Tris-Cl Buffer
1.5M Tris-Cl Stock Buffer pH 8.8
(1.5M) (X ml) = (.375M) (55 ml)

X = 13.8 ml

Water
(55 ml)–(36.7 ml + 13.8 ml) = X

X = 4.5 ml

APS
(500 µl) / (100 ml) = ( X µl) / (55 ml)

X = 275 µl

TEMED
10% the APS volume; (275 µl) / 10 = X

X = 27.5 µl

7