STEP 9: LCD DISPLAY SETUP
i. Press the SETUP button (B1) and press the SELECT button (B4)
ii. Volume control: 90, 80, 70, 60 – press CHANGE UP/DOWN to change noise level
iii. Press SELECT for LCD brightness and press CHANGE UP/DOWN to change LCD
brightness
iv. Press the SELECT button (B4)
* The following should be displayed.
SENSOR: 1
SENSOR TYPE: show the gas (example: NITROUS OXIDE)
UNITS: PSI (to change to kPA, Bar, inHg, mmHg: press button CHANGE DOWN/UP then
press SELECT)
LOW ALARM: 40 (to change: press button CHANGE DOWN/UP then press SELECT)
HIGH ALARM: 60 (to change: press button CHANGE DOWN/UP then press SELECT)
CURRENT PRESSURE: displays exact pressure for the line
CURRENT OFF SET: to re-calibrate pressure reading press button CHANGE DOWN/UP
then press SELECT
v. Repeat steps until all sensors are scanned and data is saved. Next it will show
SETUP COMPLETE.
*All gases should now be displayed. If any errors occur, repeat the steps above.
NOTE: Hold the “MUTE” button for twenty (20) seconds to display current low and high set
points.
NOTE: Press the Setup button (B1) in order to make corrections/go back.
STEP 10: CLOSING THE FRAME/MODULE ASSEMBLY
i. Close the frame panel by tightening the screws found on the frame panel to the
back box. Ensure that the screws are securely fastened to keep the LCD Alarm
closed.
ii. Carefully place the front frame over the frame panel. Screw in the screws that
were removed in Step 5, part ii. The alarm shall now be ready for use.
Installation Guide