Installation guide – Amico Alert-3 LCD Alarm User Manual

Page 11

background image

www.amico.com

11

STEP 9: LCD DISPLAY SETUP

i. Press the SETUP button (B1) and press the SELECT button (B4)

ii. Volume control: 90, 80, 70, 60 – press CHANGE UP/DOWN to change noise level

iii. Press SELECT for LCD brightness and press CHANGE UP/DOWN to change LCD

brightness

iv. Press the SELECT button (B4)

* The following should be displayed.

SENSOR: 1

SENSOR TYPE: show the gas (example: NITROUS OXIDE)

UNITS: PSI (to change to kPA, Bar, inHg, mmHg: press button CHANGE DOWN/UP then

press SELECT)

LOW ALARM: 40 (to change: press button CHANGE DOWN/UP then press SELECT)

HIGH ALARM: 60 (to change: press button CHANGE DOWN/UP then press SELECT)

CURRENT PRESSURE: displays exact pressure for the line

CURRENT OFF SET: to re-calibrate pressure reading press button CHANGE DOWN/UP

then press SELECT

v. Repeat steps until all sensors are scanned and data is saved. Next it will show

SETUP COMPLETE.

*All gases should now be displayed. If any errors occur, repeat the steps above.

NOTE: Hold the “MUTE” button for twenty (20) seconds to display current low and high set

points.

NOTE: Press the Setup button (B1) in order to make corrections/go back.

STEP 10: CLOSING THE FRAME/MODULE ASSEMBLY

i. Close the frame panel by tightening the screws found on the frame panel to the

back box. Ensure that the screws are securely fastened to keep the LCD Alarm
closed.

ii. Carefully place the front frame over the frame panel. Screw in the screws that

were removed in Step 5, part ii. The alarm shall now be ready for use.

Installation Guide