Assembly and installation mira 45 wall mount – Amico Mira 45 User Manual
Page 7
www.amico.com
7
Operating manual
(GB)
Page
16/40
6.
Assembly and Installation MIRA 45 Wall Mount
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply
voltage must correlate with the voltage noted on the fixture’s rating label. The
wall fitting is to be used exclusively with walls that are either solid, or supplied
with a suitable high-strength wall-mounting solution.
c
Drill holes in accordance with the enclosed drilling template (∅=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using
the four anchor bolts according to detailed view 2. Tap the anchor bolts
completely in, using a hammer. Tighten the nut with an open-ended
wrench (approx. 3-4turns, 23Nm torque.
Detailed view 1
d
Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the
location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and
support tube together (view 1, direction ‘b’). Align the hole in the slider with
the set-screw hole at the bottom of the support tube. Insert the set-screw
into the hole to lock the slider in place (1c). Tighten the set-screw
completely, then loosen it by a maximum 1-quarter turn and check the
operation of the arm joint.
Detailed view 2
e
Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and install the locking bolt (Detailed view
2a) from top. Tighten the screws on the anchor plate. Observe the correct
order of the washer (2b) and the serrated lock washer (2c). Now connect to
the electrical power supply, as per the wiring diagram (see sheet enclosed
with the set of bolts). Slide the ceiling cover and the O-ring to the anchor
plate.
f
Head Assembly (see page
6
8)
Mounting steps
Operating manual
(GB)
Page
16/40
6.
Assembly and Installation MIRA 45 Wall Mount
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply
voltage must correlate with the voltage noted on the fixture’s rating label. The
wall fitting is to be used exclusively with walls that are either solid, or supplied
with a suitable high-strength wall-mounting solution.
c
Drill holes in accordance with the enclosed drilling template (∅=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using
the four anchor bolts according to detailed view 2. Tap the anchor bolts
completely in, using a hammer. Tighten the nut with an open-ended
wrench (approx. 3-4turns, 23Nm torque.
Detailed view 1
d
Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the
location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and
support tube together (view 1, direction ‘b’). Align the hole in the slider with
the set-screw hole at the bottom of the support tube. Insert the set-screw
into the hole to lock the slider in place (1c). Tighten the set-screw
completely, then loosen it by a maximum 1-quarter turn and check the
operation of the arm joint.
Detailed view 2
e
Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and install the locking bolt (Detailed view
2a) from top. Tighten the screws on the anchor plate. Observe the correct
order of the washer (2b) and the serrated lock washer (2c). Now connect to
the electrical power supply, as per the wiring diagram (see sheet enclosed
with the set of bolts). Slide the ceiling cover and the O-ring to the anchor
plate.
f
Head Assembly (see page
6
8)
Mounting steps
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply voltage must
correlate with the voltage noted on the fixture’s rating label. The wall fitting is to be used
exclusively with walls that are either solid, or supplied with a suitable high-strength wall-
mounting solution.
1. Drill holes in accordance with the enclosed drilling template (Ø=8 mm; depth=55
mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts
according to detailed view 2. Tap the anchor bolts completely in, using a hammer.
Tighten the nut with an open-ended wrench (approx. 3-4 turns, 23Nm torque.
2. Perform this Step on a flat and level surface. Pull the cable through the support tube.
Insert the slider into the groove of the arm joint, noting the location of the fixing
hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1,
direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the
support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten
the set-screw completely, then loosen it by a maximum 1-quarter turn and check the
operation of the arm joint.
3. Place the O-ring and the ceiling cover over the support tube. Insert the support tube
into the anchor plate and install the locking bolt (Detailed view 2a) from top. Tighten
the screws on the anchor plate. Observe the correct order of the washer (2b) and
the serrated lock washer (2c). Now connect to the electrical power supply, as per the
wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and
the O-ring to the anchor plate.
4. Head Assembly (see page 8)
Assembly and Installation MIRA 45 Wall Mount