0 modo totalizer – Monarch Instrument PLT200 User Manual
Page 9

Display:
Dual LCD Display (5-digit upper/scrolling, 5-digit
alphanumeric lower display)
Batteries:
2 “AA” 1.5 V (DC) alkaline included
(Note: Batteries are NOT rechargeable.)
Battery Life: 30 hours continuous typical with batteries provided
External Input:
Absolute max: -0.3 V to 5 V (DC)
Minimum:
low below 1.2 V and high above 2 V (TTL compatible)
Edge:
Triggers on Positive edge
Power Out:
3.0 V nominal, approx. 2.8 V @ 20 mA max
Pulse Output: 0 V to 3.3 V (DC) pulse
Same shape as External Input signal or high when internal
optics sees a reflection
Dimensions:
6.92” (17.58 cm) H x 2.4” (6.10 cm) W x 1.6” (4.06 cm) D
Weight:
Approx. 7 oz. (210 g)
This product is designed to be safe for indoor use under the following
conditions (per IEC61010-1).
Installation Category II
per IEC 664
Pollution Degree Level II per IEC 664
Temperature: 40 °F to 105 °F (5 °C to 40 °C)
Humidity:
Maximum relative humidity of 80% for temperatures up
to 88 °F (31 °C) decreasing linearly to 50% relative
humidity at 100 °F (40 °C). Humidity non-condensing.
Specifications subject to change without notice.
E-6
S-17
1.
Encender la
unidad
2.
Entre modo
Setup
3.
Entre
selección de
modo
4.
Seleccione
Modo
TOTAL
5.
Guardar y
avanzar
O
Repita hasta
TOTAL es
despliegado
El último modo
seleccionado se
despliega
1a. Para cambiar entre
modo Fijo y no Fijo
Differentes mensajes se despliegan
para operatión Interna o Externa
Locked On
ROS Interno o Externo:
RCA Externo:
EXTRN, luego mensaje
avanza, luego últimas
unidades seleccionadas
Ultimas
unidades
seleccionadas
Mantenga presionado el boton
10.0 Modo TOTALizer
10.1 Arreglo Para Modo TOTALizer