DPD CHLORINE

TEST METHOD SPECIFICATIONS

APPLICATION

Drinking water supplies and distribution systems, swimming pool and spas, sewage
and chlorinated waste waters, process waters and sanitizing solutions.

RANGE

0 to 4.0 ppm Chlorine (may be extended by dilution)

METHOD

In the absence of Iodide, Free Available Chlorine reacts instantly with the buffered
diethyl-p-phenylenediamine indicator (DPD) to produce a red color in proportion to
the amount of chlorine present. Subsequent addition of potassium iodide produces a
rapid color response from the combined forms of chlorine (chloramines).

HANDLING & PRESERVATION

Chlorine in aqueous solutions, particularly weak solutions, is not stable. Exposure to
sunlight or agitation will accelerate the reduction of chlorine. Fill sample containers to
the top and cap tightly. Analyze samples as soon as possible after collection.

INTERFERENCES

The only interfering substance likely to be encountered is oxidized manganese. The
extent of this interference can be determined by treating a sample with sodium arsenite
to destroy the chlorine present, so that the amount of interference can be measured.

CALIBRATION

The single test colorimeter is precalibrated. In order to comply with NPDWR or
NPDES reporting regulations, the calibration should be checked periodically by
using a set of reference standards including a 0 mg/L blank and 0.3, 1.0, and 3.5
mg/L chlorine. To prepare these standards, a LaMotte 1000 mg/L standard chlorine
equivalent solution (Code 3858) is available. Consult with your local regulatory
agency to determine.